> For the complete documentation index, see [llms.txt](https://docs.aethlabs.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://docs.aethlabs.com/microaethalometer-tm-ma450/quick-start-guide/04-installation-for-outdoor-deployment.md).

# Installation for Outdoor Deployment

For detailed installation instructions, see the [Installation chapter](https://github.com/AethLabs/aethlabs-manuals/tree/main/gitbook/microaethalometer-tm-ma450/manual/installation.md).

{% stepper %}
{% step %}

### Pick installation location

It is always recommended to shield the instrument from direct solar radiation. The inlet port, outlet port, and all connections must be properly protected from the environment.
{% endstep %}

{% step %}

### Mount the instrument

Mount the MA450 using the rear sun shield mounting locations
{% endstep %}

{% step %}

### Install inlet protection or sampling hose

Install the included MA450 Inlet Protection Kit or [Inlet Protection Kit (WPK-20-25)](https://github.com/AethLabs/aethlabs-manuals/tree/main/gitbook/accessories/inlet-protection-kit.md) or other appropriate inlet sampling train to protect against water intrusion, insects, and in-line condensation.
{% endstep %}

{% step %}

### Route hose(s)

Carefully route all inlet and outlet hoses to prevent water from entering the ports.
{% endstep %}

{% step %}

### Connect power

Connect the supplied 12V outdoor-rated power adapter to the sealed M8 circular connector labeled **12V DC IN**. If the optional built-in Nafion dryer is installed, also connect the second 12V supply to the connector labeled **DRYER 12V DC IN**. See [Powering the MA450](/microaethalometer-tm-ma450/quick-start-guide/03-powering-the-ma450.md).
{% endstep %}
{% endstepper %}

{% hint style="warning" %}
Be sure that the aerosol inlet to the microAethalometer™ is at atmospheric pressure and that there is no differential pressure between the aerosol inlet and the operating environment. If sampling from a plenum or chamber, it must be at the same pressure as the instrument. If other instrumentation is measuring in parallel, always use a dedicated sampling line for the microAethalometer™.
{% endhint %}

| <img src="https://272274374-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2F1fOQ2MnLWcq0ntHW3LLY%2Fuploads%2Fgit-blob-a93db31457864152212fe12658bf37d213f106ef%2Fma450-outdoor-enclosure-inlet-protection.jpg?alt=media" alt="Front view of the enclosure with the Inlet Protection Kit connected." data-size="original"> | <img src="https://272274374-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2F1fOQ2MnLWcq0ntHW3LLY%2Fuploads%2Fgit-blob-767fdc3d3e71eccc8eecd03c39bc11b14cb691bb%2Fma450-outdoor-enclosure-inlet-protection-ports.jpg?alt=media" alt="Connector-panel detail: water/bug trap, rain-shield hood, and inlet hose routing." data-size="original"> |
| --------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- | ---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- |